Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 12B: (1 year) CAL VER /5700 INSTRUMENT: Fluke 18: (1 year) CAL VER /5700 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 187 CONFIGURATION: Fluke 5700A STANDARD: Capacitance Decade ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_18_cw.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Capacitance Decade 1.002 ASK- R N P F V 1.003 ASK+ K 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP No warm-up time requirement. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 90%. 1.007 DISP Set the Rotary Switch to the OFF position. 1.007 DISP 1.007 DISP Note: Erroneous result may occur if slide switch 1.007 DISP is not placed in OFF position before beginning tests. 1.008 HEAD {DC VOLTAGE TESTS}: 4V Range 1.009 PIC 57_18_cw 1.010 DISP Set the Rotary Switch to the center position. 1.010 DISP Press RANGE to select the 4V Range (0.000). 1.011 ASK- Q 1.012 5700 0.000V S 2W 1.013 HEAD DC VOLTAGE TESTS: {4V Range (MAX)} 1.014 DISP Press MIN/MAX to select Min/Max Mode. 1.014 DISP Press MIN/MAX again to display Max. 1.015 MEMI Enter UUT reading in volts: 1.016 MEME 1.017 MEMC 4 V 0.012U 2.001 MEME 2.002 HEAD DC VOLTAGE TESTS: {4V Range (MIN)} 2.003 DISP Press MIN/MAX to display Min. 2.004 MEMI Enter UUT reading in volts: 2.005 MEME 2.006 MEMC 4 V 0.012U 3.001 DISP Press MIN/MAX for 2s to exit Min/Max Mode. 3.002 ASK+ Q 3.003 HEAD DC VOLTAGE TESTS: {4V Range (LoZ)} 3.004 DISP Set the Rotary Switch to the right position. 3.005 5700 4 0.000V 0.002U 2W 4.001 HEAD DC VOLTAGE TESTS: {40V Range (LoZ)} 4.002 DISP Press RANGE to select the 40V Range (0.00). 4.003 5700 40 0.02V 0.01U 2W 5.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 5.002 DISP Set the Rotary Switch to the center position. 5.003 DISP Press RANGE to select the 400V Range (0.0). 5.004 5700 400 -0.2V 0.1U 2W 6.001 HEAD DC VOLTAGE TESTS: {4V Range (HiZ)} 6.002 DISP Press RANGE for 2s to select autorange. 6.003 5700 4 3.500V 0.034U 2W 7.001 HEAD DC VOLTAGE TESTS: {40V Range (HiZ)} 7.002 5700 40 -35.00V 0.33U 2W 8.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 8.002 5700 400 100.0V 1.0U 2W 9.001 HEAD DC VOLTAGE TESTS: {600V Range (HiZ)} 9.002 5700 600 -600V 6U 2W 10.001 HEAD {AC VOLTAGE TESTS} 10.002 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 10.003 DISP Press the Function Button to select VAC. 10.003 DISP Press RANGE to select the 4V range (0.000). 10.004 5700 0.000V S 2W 10.005 MEMI Enter UUT reading in volts: 10.006 MEME 10.007 MEMC 4 V +0.003U 11.001 ASK- Q 11.002 5700 0.000V S 2W 11.003 HEAD AC VOLTAGE TESTS: {4V Range (MAX)} 11.004 DISP Press MIN/MAX to select Min/Max Mode. 11.004 DISP Press MIN/MAX again to display Max. 11.005 MEMI Enter UUT reading in volts: 11.006 MEME 11.007 MEMC 4 V +0.040U 12.001 MEME 12.002 HEAD AC VOLTAGE TESTS: {4V Range (MIN)} 12.003 DISP Press MIN/MAX to display Min. 12.004 MEMI Enter UUT reading in volts: 12.005 MEME # Lower limit should be 0.000 but firmware bug in early units allowed -0.004 12.006 MEMC 4 V -.004U +.040U 13.001 DISP Press MIN/MAX for 2s to exit Min/Max Mode. 13.002 ASK+ Q 13.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 13.004 5700 4 0.010V 0.003U 50H O 2W 14.001 5700 4 0.010V 0.003U 400H O 2W 15.001 HEAD AC VOLTAGE TESTS: {4V Range (LoZ)} 15.002 DISP Set the Rotary Switch to the right position. 15.003 5700 4 3.500V 0.070U 50H 2W 16.001 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 16.002 DISP Set the Rotary Switch to the center position. 16.003 5700 4 3.500V 0.070U 400H 2W 17.001 HEAD AC VOLTAGE TESTS: {40V Range (HiZ)} 17.002 DISP Press RANGE to select the 40V range (0.00). 17.003 5700 40 35.00V 0.70U 50H 2W 18.001 DISP Set the Rotary Switch to the right position. 18.002 HEAD AC VOLTAGE TESTS: {40V Range (LoZ)} 18.003 5700 40 35.00V 0.70U 400H 2W 19.001 HEAD AC VOLTAGE TESTS: {400V Range (LoZ)} 19.002 DISP Press RANGE to select the 400V range (0.0). # Warmup 2k resistor so 350V will not current limit 5700 19.003 5700 100.0V 80H S 2W 19.004 5700 150.0V 80H S 2W 19.005 5700 200.0V 80H S 2W 19.006 5700 250.0V 80H S 2W 19.007 5700 300.0V 80H S 2W 19.008 5700 400 350.0V 7.0U 80H 2W 20.001 HEAD AC VOLTAGE TESTS: {600V Range (HiZ)} 20.002 DISP Set the Rotary Switch to the center position. 20.002 DISP Press RANGE to select the 600V Range (0). 20.003 5700 600 600V 14U 400H 2W 21.001 HEAD {CONTINUITY TEST} 21.002 DISP Set the Rotary Switch to the right position. 21.002 DISP Press RANGE for 2s to select autorange. 21.003 5700 2.000V 0.020U 2W 22.001 HEAD CONTINUITY TEST: {VCHEK} 22.002 5700 3.430V 0.033U 2W 23.001 5700 -0.500V 0.007U 2W 24.001 HEAD {RESISTANCE TESTS} 24.002 DISP Press the Function Button to select Ohms. 24.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 24.004 5700 400 0.0Z +0.2U CW 25.001 5700 400 1.0Z 0.2U CW 26.001 5700 400 190.0Z 0.9% 0.2U CW 27.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 27.002 5700 4 1.000kZ 0.010U CW 28.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 28.002 5700 40 19.00kZ 0.9% 0.01U CW 29.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 29.002 5700 400 100.0kZ 1.0U 2W 30.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 30.002 5700 4 1.900MZ 0.9% 0.001U 2W 31.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 31.002 5700 40 19.00MZ 1.5% 0.03U 2W 32.001 HEAD RESISTANCE TESTS: {VCHEK} 32.002 5700 1.960V 0.020U 2W 33.001 DISP Disconnect the 5700A from the UUT. 33.002 HEAD {CAPACITANCE TESTS} 33.003 DISP Press the Capacitance Button. 33.004 HEAD CAPACITANCE TESTS: {1uF Range} 33.005 MATH MEM1 = 0.0 33.006 MEMI Enter UUT reading in microfarads: 33.007 MEME 33.008 MEMC 1 uF 0.001U 34.001 DISP Connect the capacitance decade to the UUT. 34.002 DISP Set the decade to 0.950uF. 34.003 ACC 0.950uF 0.5% 34.004 MEMI Enter UUT reading in microfarads: 34.005 MEME 34.006 MEMC uF 0.020U 35.001 END